Ohio State is in the process of revising websites and program materials to accurately reflect compliance with the law. While this work occurs, language referencing protected class status or other activities prohibited by Ohio Senate Bill 1 may still appear in some places. However, all programs and activities are being administered in compliance with federal and state law.

Main navigation

Facts About Non-Mercury Thermometers

Topic:

Thermometer Storage and Fluid Separation

The science and development of non-mercury thermometers have made great improvements over the past few years. The blue spirit thermometers listed contain non-toxic isoamyl benzoate and dye. These thermometers can be stored horizontally; their separation rate is equal to or better than mercury thermometers. The red spirit thermometers that contain pentane or xylene and dye (-100° to 50°C and – 50° to 50°C) should be stored vertically to reduce the chances of separation.

Temperature Ranges and Limits

There are non-mercury thermometers with temperature and accuracy ranges equal to most mercury thermometers. An exception is on the high end of the temperature scale. Until recently 260° C was the typical limit for non-mercury thermometers. A new Brown-Spirit thermometer has a maximum range greater than 300° C but the accuracy limit can be ± 5°C at high temperatures. If scientific procedures cannot be performed with non-mercury thermometers, Teflon coated mercury thermometers should be used. The Teflon coating will help to reduce releases of mercury should a thermometer be broken.

Scale Divisions and Accuracy Limits

Non-mercury thermometers have scale divisions equal to mercury thermometers. Most range from 0.5° to 1° C. Accuracy limits for non-mercury thermometers below 150° C are also compatible with mercury thermometers; accuracy is typically ± 1° - 2° C. For thermometers that have a high range of 260° C, accuracy below 100° C is ± 1°-2° C for both mercury and non-mercury thermometers. Above 100° C mercury thermometers have an accuracy range of ± 1.5° C, while non-mercury have an accuracy limit of ± 3° C.

National Institute of Standards and Technology

All of the non-mercury thermometers listed meet NIST standards.

Cleanup and Disposal

The non-mercury thermometers are non-toxic and environmentally safe. The broken glass should be placed in a sharps container to prevent injury. The remaining liquid can be cleaned up with soap and water.